Radnoti 酸素チャンバー

Radnoti 酸素チャンバーは、灌流回路内の溶液の温度を保持しながら酸素を注入するパーツです。

概観

Radnoti 酸素チャンバーは、温度を保持しながら灌流回路内の溶液へ酸素を注入する装置です。本製品はホウケイ酸ガラス製で、ウォータージャケットの外側コンパートメントを含み、Radnoti Quick Disconnectフィッティングを使用して温水灌流回路へ取り付けます。さらに、シングルリングクランプユニバーサルスタンドクランプを使用してRadnotiラボスタンドに取り付けることができます。

Product Configurations

Radnoti Oxygenating Chamber (Sheet Flow)

製品コード
120144

Radnoti Oxygenating Chamber (Sheet Flow) 

An oxygenating chamber that is designed to be vertically attached in a Radnoti Working Heart apparatus for rat hearts. Its upper chamber is a condenser, which is formed from six linked spheres that provide the maximal surface area for rewarming and reoxygenation of the perfusate. Perfusate passing through the condenser is collected in its lower chamber, which is designed to collect the perfusate. It has a port that allows the insertion of an Oxygenating Tube with Teflon Needle Valve to provide controlled gas delivery. Additionally, it has another specialized port for the option of inserting either a temperature or pH probe to monitor temperature or pH.

This oxygenating chamber has seven Luer ports, one entry port at the top, three side ports near the top of the upper chamber, two side ports connecting to the lower chamber, and one exit port at the bottom. Attachment of either 1- or 3-Way Luer Taps allows the top and side ports to be either used for entry or exit of perfusate overflow, perfusate return (for working heart mode), gassing, perfusate sampling or addition of drugs. The bottom exit port permits the reoxygenated perfusate to be pumped back to the isolated heart. It has a 160 mL volume, 370 mm overall average height and 48 mm water jacket OD.


 

Image
Sheet flow oxygenating chamber
$1,200.00
Ships from supplier Estimated delivery 1 month

Radnoti Oxygenating Chamber (Membrane)

製品コード
130144

Radnoti Oxygenating Chamber (Membrane) 

An oxygenating chamber that is designed to be horizontally attached to a Radnoti Lab Stand via two Single Ring Clamps, which is then used to gas solutions that are being perfused at very low flow rates required for certain organs, tissues, and cells. They include Isolated Rodent Liver/Kidney and Mouse Langendorff and Working Heart setups.

The oxygenating chamber has a lid that seals an inner incubating chamber, and has gas inflow and outflow ports, which allow the chamber to be gassed from a gas supply (depending on the user and laboratory facilities). The gas pressure should not be too high, but within the recommended range, which is similar to a gentle stream when played over the hand. Attached to the lid is a three rod cylindrical frame that is wrapped by a coil of silicone tubing, which is highly permeable to oxygen, carbon dioxide and most other common gasses. Therefore, firstly the chamber should be allowed to come to temperature, before passing the perfusate through the silicone tubing, which permits rapid gas equilibration with most gas mixtures.  
   
 


 

Image
Membrane Oxygenating Chamber

This chamber was designed to accommodate a range of tubing lengths, wall thickness and diameters, which will affect the flow rate (effective gas exchange rate) and hence the oxygenation level. Further information on gassing and silicone tubing is available under Details and Tech Specs.

It is good practice to clean the oxygenator prior to use. The glass and polypropylene cap, and Luer valves and silicone tubing can be cleaned with water and soap or general glassware cleaner, then rinsed thoroughly with distilled water. The silicone tubing coil should be flushed copiously with water (1-2 liters). Additionally, if required the borosilicate glass oxygenator with thermally stable polypropylene parts can be sterilized by autoclaving or ethylene gas. Note: Use of certain materials that adhere to silicone tubing may require tubing changes to avoid contamination.

$1,200.00
Ships from supplier Estimated delivery 1 month

Lab Builder – Find the right products for your research.

1Select your research area

2Use our smart guide to tell us your needs

3Get instant product recommendations

    Try Lab Builder

    サポート

    サポート記事

    Our Support Articles contain answers to common questions, how-tos, solutions, and documentation.

    詳細と技術仕様

    Technical

    Radnoti酸素チャンバー(膜)のガス抜き 
    標準的なガスボンベでガスを供給する場合は、正確な制御が可能なセカンドステージバルブ制御装置が必要です。大型モデルの場合、ガス管は蓋アセンブリの下側にあるリリーフバルブアクセスに接触させることができます。ガス圧が高すぎるとリリーフバルブが作動します。ガスは流入チェックバルブアセンブリーからインキュベーターへ入り、流出チェックバルブアセンブリーから排出されます。適切なガス圧が設定されたら、ガスラインを流入逆止弁アセンブリに接続します。

    Radnoti酸素チャンバー(膜)用シリコンチューブ
    異なるサイズや長さのシリコンチューブを選択することで、チャンバーの内部容積を調整することができます。チューブを通過する流体の酸素濃度はチューブの長さと流体がチャンバー内に滞留する時間に直接依存し、表面積/容積比に反比例します。

    シリコンチューブごとの酸素濃度参考表

    チューブサイズチューブの長さ(inch)流量 (mL/min)酸素化率内部容量 (mL)
    0.078 x 0.125

    50′

    80

    > 95%

    50

    0.078 x 0.125

    25′

    35

    > 90%

    25

    0.078 x 0.125

    25′

    80

    > 78%

    25

    0.132 x 0.183

    50′

    48

    > 95%

    134

    実際の酸素濃度は上記表を目安にして検証してください。溶液の入口/出口に電極を設置し、張力と流量を設定することで酸素濃度の測定ができます。

    製品コード

    120144
    Radnoti Oxygenating Chamber (Sheet Flow)
    130144
    Radnoti Oxygenating Chamber (Membrane)